From: Performances of domiciliary ventilators compared by using a parametric procedure
Restrictive model | Obstructive model | ||||||||
---|---|---|---|---|---|---|---|---|---|
η h | τ pr | η l | ε | η h | τ pr | η l | ε | ||
BiPAP A40 | 5 l/min | 1 | 25 % | 82 % | 1 l/min | 1 | 75 % | 97 % | |
Elisée 150 | 1 | 4 | 25 % | 80 % | 1 | 2 | 75 % | 94 % | |
Monnal T50 | 5 l/min | 1 | 25 % | 71 % | 0.5 l/min | 4 | 75 % | 37 % | |
S9 VPAP ST | Very low | Min | Very low | 69 % | Very high | Min | Very high | 88 % | |
Smartair ST | 3 | 1 | –75 % | 74 % | 1 | 1 | –25 % | 71 % | |
SOMNOvent | 1 | Std | — | 92 % | –1 | Std | — | 73 % | |
Stellar 100 | Very low | MIN | Very low | 72 % | Very high | MIN | Very high | 89 % | |
Trilogy 100 | 5 l/min | 1 | 25 % | 83 % | 1 l/min | 1 | 75 % | 96 % | |
Vivo 50 | 1 | 9 | 7 | 100 % | 1 | 1 | 3 | 82 % |