Table 5 Values of the two parameters (high pressure triggering sensitivity η h and pressure rise duration τ pr) maximizing the synchronizability ε according to the pulmonary model considered. These values were obtained from the tests performed under the protocol we defined and do not imply that a better combination of settings can not be found. The pressure release triggering sensitivity η l is also reported

From: Performances of domiciliary ventilators compared by using a parametric procedure

  Restrictive model   Obstructive model
  η h τ pr η l ε   η h τ pr η l ε
BiPAP A40 5 l/min 1 25 % 82 %   1 l/min 1 75 % 97 %
Elisée 150 1 4 25 % 80 %   1 2 75 % 94 %
Monnal T50 5 l/min 1 25 % 71 %   0.5 l/min 4 75 % 37 %
S9 VPAP ST Very low Min Very low 69 %   Very high Min Very high 88 %
Smartair ST 3 1 –75 % 74 %   1 1 –25 % 71 %
SOMNOvent 1 Std 92 %   –1 Std 73 %
Stellar 100 Very low MIN Very low 72 %   Very high MIN Very high 89 %
Trilogy 100 5 l/min 1 25 % 83 %   1 l/min 1 75 % 96 %
Vivo 50 1 9 7 100 %   1 1 3 82 %